Autoclaving liquids presents some unique challenges. Sterilization cycles for liquid loads take a longer time to complete since liquids have a high heat capacity. As opposed to solids, liquids take a lot longer to heat up and cool down, and the total cycle time is increased dramatically as a result. When autoclaving liquids, certain cycle features can be incorporated into the sterilization process to optimize its efficiency and sterilization effectiveness.
Introduction:
Autoclaving liquids presents some unique challenges. Sterilization cycles for liquid loads take a longer time to complete since liquids have a high heat capacity. As opposed to solids, liquids take a lot longer to heat up and cool down, and the total cycle time is increased dramatically as a result. When autoclaving liquids, certain cycle features can be incorporated into the sterilization process to optimize its efficiency and sterilization effectiveness.
Challenges of Sterilizing Liquids:
When sterilizing solids, air removal from the chamber is usually performed to allow steam which is the sterilization medium to penetrate the load and make contact with all surfaces allowing efficient sterilization. However, when sterilizing liquids, steam is used as the medium to heat up the liquid. Sterilization takes place when the liquid itself reaches the desired sterilization temperature and is held at a certain length of time to achieve sterility.
Reducing the boil-over effect is another challenge faced for liquid loads undergoing a sterilization process. Boil-over effect can result in in liquid spillage from the vessel and even glassware breakage if it occurs violently. This can cause a significant loss of liquid volume as well as creating a mess in the autoclave chamber. This is an undesirable situation and can potentially be dangerous for the user if the liquid is biohazardous or hazardous. Due to the delicate nature of rapidly heating and cooling liquids under pressure, special precautions must be taken to avoid this occurrence from happening.
Liquid Cooling
Reducing the cooling times for liquid sterilization helps not only to save time but also protect the integrity of the load and ensure the liquids are not “overcooked”. There are two cooling applications available for liquid cooling:
Fast Liquid Cooling: Cools down the load up to 75% faster than normal cooling under ambient (regular room temperature) conditions.
Upon completion of sterilization phase, chamber pressure is increased by forcing compressed air through a microbiological filter into the chamber. This increase in pressure prevents boil-over, spills and cracks in containers that would normally occur under low-pressure conditions. Cold water is circulated through the chamber jacket to cool the chamber and. This combination of introducing cold water and pressurized air helps reduce the temperature of the liquid load quickly and safely.
Super Fast Liquid Cooling: Super fast liquid cooling can reduce liquid cycle times by as much as 90%.
In addition to the introduction of high-pressure air inside the chamber and cold water circulating in the walls of the chamber, a fan can be used to further accelerate cooling by speeding up the circulation of air, which transfers the heat from the load to the cold chamber walls more efficiently.
The solid red line represents temperature for fast cooling, and the dotted red line represents temperature for standard cooling. The solid blue line represents pressure for fast cooling, and the dotted blue line represents pressure for standard cooling
Two PT 100 Temperature Sensors
The chamber temperature heats and cools faster compared to the liquid inside the container. To account for this disparity, two flexible temperature sensors can be placed inside the liquid load, giving an accurate reading of the temperature of the liquid. This ensures that the liquid’s temperature reaches the proper level for sterilization. The two flexible temperature sensors are also used to control the conditions inside the chamber during the cooling phase. This helps to avoid dangerous messes caused by boil-over and ensure safe conditions for opening the door.
Using two temperature sensors (in different vessels) ensures user safety. In the event of a container breakage, there will be a difference between the 2 readings prompting the autoclave operator to halt the cycle and keep the door locked until the temperature and pressure inside the chamber return to safe conditions for opening the door.
F0 Feature for Heat-Sensitive Liquid Media
Laboratory autoclaves are designed to be able to perform the F0 feature (pronounced F-zero). This is a feature that reduces exposure time of the load to high temperatures, making it suitable for delicate heat-sensitive liquid media. In minimizing the overall time that liquid media are exposed to high temperatures, the integrity of the load is maintained, as well as reducing overall cycle time and energy costs for the lab. It allows the autoclave operator to consider the heat energy contributed during the heat-up time as contributing to sterilization. This is calculated by empirically tested tables of F0 values.
The F0 feature is useful for a very delicate liquid, which can be easily caramelized and therefore disqualified for use. The autoclave operator can calculate a lower cycle time for sterilizing the liquid based on the F0 empirical tables. If a normal sterilization holding time is set to 15 minutes, it may be reduced by seconds/minutes based on the F0 calculation. This will help prevent overcooking the liquid while still ensuring proper sterilization.
In conclusion, the process for sterilizing liquids in an autoclave is different from the usual process. However, with the incorporation of certain features such as liquid cooling, use of temperature sensors and F0 cycles, this will help to ensure a safe and efficient sterilization process.
The information below is required for social login
Sign In
Create New Account
Autoclaving liquids presents some unique challenges. Sterilization cycles for liquid loads take a longer time to complete since liquids have a high heat capacity. As opposed to solids, liquids take a lot longer to heat up and cool down, and the total cycle time is increased dramatically as a result. When autoclaving liquids, certain cycle features can be incorporated into the sterilization process to optimize its efficiency and sterilization effectiveness.
Introduction:
Autoclaving liquids presents some unique challenges. Sterilization cycles for liquid loads take a longer time to complete since liquids have a high heat capacity. As opposed to solids, liquids take a lot longer to heat up and cool down, and the total cycle time is increased dramatically as a result. When autoclaving liquids, certain cycle features can be incorporated into the sterilization process to optimize its efficiency and sterilization effectiveness.
Challenges of Sterilizing Liquids:
When sterilizing solids, air removal from the chamber is usually performed to allow steam which is the sterilization medium to penetrate the load and make contact with all surfaces allowing efficient sterilization. However, when sterilizing liquids, steam is used as the medium to heat up the liquid. Sterilization takes place when the liquid itself reaches the desired sterilization temperature and is held at a certain length of time to achieve sterility.
Reducing the boil-over effect is another challenge faced for liquid loads undergoing a sterilization process. Boil-over effect can result in in liquid spillage from the vessel and even glassware breakage if it occurs violently. This can cause a significant loss of liquid volume as well as creating a mess in the autoclave chamber. This is an undesirable situation and can potentially be dangerous for the user if the liquid is biohazardous or hazardous. Due to the delicate nature of rapidly heating and cooling liquids under pressure, special precautions must be taken to avoid this occurrence from happening.
Liquid Cooling
Reducing the cooling times for liquid sterilization helps not only to save time but also protect the integrity of the load and ensure the liquids are not “overcooked”. There are two cooling applications available for liquid cooling:
Upon completion of sterilization phase, chamber pressure is increased by forcing compressed air through a microbiological filter into the chamber. This increase in pressure prevents boil-over, spills and cracks in containers that would normally occur under low-pressure conditions. Cold water is circulated through the chamber jacket to cool the chamber and. This combination of introducing cold water and pressurized air helps reduce the temperature of the liquid load quickly and safely.
In addition to the introduction of high-pressure air inside the chamber and cold water circulating in the walls of the chamber, a fan can be used to further accelerate cooling by speeding up the circulation of air, which transfers the heat from the load to the cold chamber walls more efficiently.
The solid red line represents temperature for fast cooling, and the dotted red line represents temperature for standard cooling. The solid blue line represents pressure for fast cooling, and the dotted blue line represents pressure for standard cooling
Two PT 100 Temperature Sensors
The chamber temperature heats and cools faster compared to the liquid inside the container. To account for this disparity, two flexible temperature sensors can be placed inside the liquid load, giving an accurate reading of the temperature of the liquid. This ensures that the liquid’s temperature reaches the proper level for sterilization. The two flexible temperature sensors are also used to control the conditions inside the chamber during the cooling phase. This helps to avoid dangerous messes caused by boil-over and ensure safe conditions for opening the door.
Using two temperature sensors (in different vessels) ensures user safety. In the event of a container breakage, there will be a difference between the 2 readings prompting the autoclave operator to halt the cycle and keep the door locked until the temperature and pressure inside the chamber return to safe conditions for opening the door.
F0 Feature for Heat-Sensitive Liquid Media
Laboratory autoclaves are designed to be able to perform the F0 feature (pronounced F-zero). This is a feature that reduces exposure time of the load to high temperatures, making it suitable for delicate heat-sensitive liquid media. In minimizing the overall time that liquid media are exposed to high temperatures, the integrity of the load is maintained, as well as reducing overall cycle time and energy costs for the lab. It allows the autoclave operator to consider the heat energy contributed during the heat-up time as contributing to sterilization. This is calculated by empirically tested tables of F0 values.
The F0 feature is useful for a very delicate liquid, which can be easily caramelized and therefore disqualified for use. The autoclave operator can calculate a lower cycle time for sterilizing the liquid based on the F0 empirical tables. If a normal sterilization holding time is set to 15 minutes, it may be reduced by seconds/minutes based on the F0 calculation. This will help prevent overcooking the liquid while still ensuring proper sterilization.
In conclusion, the process for sterilizing liquids in an autoclave is different from the usual process. However, with the incorporation of certain features such as liquid cooling, use of temperature sensors and F0 cycles, this will help to ensure a safe and efficient sterilization process.